ASL Calibration¶
- Widgets -> ASL Calibration
The ASL Calibration widget is required to turn the perfusion images from ASL model fitting into physical units (ml/100g/min).
Basic data specification¶
To do calibration you specify a data set containing a perfusion image. If you want to calibrate an image containing perfusion
variance, the Data type
selection must be changed (because the scaling factors must be squared in this case).
Calibration methods¶
Two calibration methods are provided:
Voxelwise
calibration, in which an M0 correction factor is determined for each voxel from the calibration image.Reference region
calibration, in which a single M0 correction factor is determined for the whole image, by analysing a region of the data containing a single tissue type (typically CSF).
Calibration data¶
Certain parameters are required regardless of which calibration method you use:
Voxelwise calibration¶
Voxelwise calibration requires a generic estimate of the T1 and the partition coefficient which will be applied to all voxels.
Reference region calibration¶
The reference region method requires a Reference ROI
which identifies a particular tissue type. This would normally be
created by a segmentation tool such as FAST, however you could also use the ROI Builder
to identify a region of the
calibration image of a known tissue type. CSF is the most common.
The T1, T2 and partition coefficient for this tissue type must be specified. Default values are provided for CSF, WM and GM. In addition the sequence TE and blood T1 estimates are required.
Output¶
Calibration returns a new data set with the suffix _calib
, for example perfusion_calib
. In the case of reference region calibration, the calibrated image will simply be a scaled copy of the original perfusion image - as shown if we use the Compare Data
widget:
In voxelwise calibration this is not the case although there should still be an approximate linear relationship in the areas of interest - the comparison below is within a brain mask: